| Specifications |
| General
Functions |
| Output Range |
| Nominal |
Min |
Rated |
Max |
| +5V |
0.2A |
10A |
16A |
| +12V |
0A |
6A |
8A |
| -12V |
0A |
0.5A |
1A |
| +5Vsb |
0A |
0.5A |
1A |
|
| Input
Voltage |
85~265V AC
|
| Ripple and Noise |
Less than 1% of each output at rated load |
| Hold-up
Time |
20ms min |
Efficiency
|
Higher than 82% |
| Leakage
current |
Less than 300�gA at 264V AC |
| Protection |
Over voltage/Load protection and short circuit protection |
| Altitude |
0~10,000 ft |
| TTL signal compatible
power good signal |
| MTBF |
-ACE-713APM: 143,781 hours |
| Vibration |
10-55Hz at 2G 3min. period, 30 min. along X, Y and
Z axis |
| Shock |
10G for 11ms once on each X,Y and Z axis |
| Safety |
UL 2601-1 , IEC 601-1 , EN 60601-1 |
| EMI |
FCC docket 20780 curve "B", EN 55011 class "B", EN 61000-3-2
class "D" |
| EMC |
-EN 61000-4-2 4KV contacto F8KV air discharge criterion
B
-EN 61000-4-3 10V/M with 80% AM criterion A
-EN 61000-4-4 2KV criterion B
-EN 61000-4-5 Line to Line 1KV ; Line to ground 2KV criterion
B
-EN 61000-4-6 10V with 80% AM criterion A
-EN 61000-4-8 30A/m criterion A
-EN 61000-4-11 30% dips 10ms , criterion B
60% dips 100ms , criterion C >95% dips 5000ms , criterion
C |
| Environmental and
Mechanical |
| Opeating Temperature |
0~70�XC (Derating curve will occur 50�XC~70�XC) |
| Storage Temperature |
-40�XC~75�XC
|
| Dimensions |
89(W)x39(H)x152.4(D)mm |
| ACE-713APM |
130W Medical ATX Power Supply with PFC function |